Description
Rotating anode X-ray tube housing assembly for the purpose of general diagnostic X-ray procedures.
Specially processed Rhenium-tungsten faced molybdenum target of 74 mm diameter.
These tubes have foci 2.0 mm and 1.0 mm, and are available for a maximum tube voltage 125 kV with Single-phase or Three-phase generator.
Accommodated with IEC 60526 type high-voltage cable receptacles.
General Data:
IEC Classification: Class I
Electrical:
Circuit (Center-grounded): Single-phase full-wave rectified or Three-phase full-wave rectified
Operating Tube Voltage:
Radiographic: 40 ~ 125 kV Max.
Fluoroscopic: 40 ~ 125 kV Max.
Focal Spot:
Large Focus: 2.0 mm
Small Focus: 1.0 mm
Input Energy (at 0.1s):
50 Hz | 60 Hz | |
Large Focus: | 42.5 Kw | 47 kW |
Small Focus: | 21 kW | 22.5 kW |
Motor Ratings:
Duty | Starting | Running | |
Power source (Hz) | 50/60 | 50/60 | |
Input power (W) | 1050 | 270 | 43 |
Voltage (V) | 200 | 100 | 40 |
Current (A) | 6.0 | 3.0 | 1.2 |
Min. Speed up (s) | 0.8 | 1.5 | – |
Capacitor (μF) | 24 | 24 | 24 |
Anode Speed:
50 Hz – 2700 min-1 Min.
60 Hz – 3200 min-1 Min.
Resistance between Housing and Low Voltage Terminals – 2 MΩ Min.
Mechanical:
Dimensions: See dimensional outline
Overall Length: 479 mm
Maximum Diameter: 152.4 mm
Target:
Angle: 16 degrees
Construction: Rhenium-Tungsten-faced molybdenum
Permanent Filtration: 0.9 mm Al / 75 kV IEC 60522/1999
Radiation Protection (To meet the requirements of IEC 60601-1-3):
Leakage Technique Factor – 125 kV 4 mA
X-ray Coverage: 354 × 354 mm at SID 750 mm
Weight (Approx.): 16 kg
High Tension Terminals: To meet the requirements of IEC 60526
Cooling Method: Natural or forced air
Tube Housing Model Number:
E7239X: XH-121
E7239FX: XH-126
E7239GX: XH-150